Choosing the Right Mass Flow Controller for Your Application


There are four main considerations to evaluate in order to specify a Thermal Mass Flow Controller (MFC) for your application. Flow range, gas conditions, desired accuracy, and  communications are all important constraints. There are a five additional considerations including interior finish for high and ultra high purity applications, materials used in the flow path, the environment such as installations near furnaces or some sort of NEMA rating, mounting attitude such as vertical flow up, and serviceability.



MFCs are used to measure and control the flow of gasses in a process. The overall flow rate is an important consideration. Some processes require very small flows usually specified in standard cubic centimeters per minute (SCCM). Higher flows are typically given in standard liters per minute (SLPM) or standard cubic feet per minute (SCFM). Many manufacturers make devices able to measure flows of 200 sccm up to 30 slpm. The same body size can be used for this wide range of flows; however, the internal components are changed to give a more precise flow range. Typically, once you choose a maximum flow, you are able to measure and control down from there a certain amount. Having a 10 to 1 turndown is common. Being able to control to a 50:1 or even 100:1 is available from some of the leading manufactures. Setting the maximum flow to something very small like 10 sccm or even 3 sccm and then being able to measure and control from that point down to 2% is remarkable, but is standard for some. Once the desired flow is more than 30 slpm (or 1 scfm), usually the physical size of the MFC body changes a bit. This allows greater flow as there are larger passageways inside the device. Flows to 100 slpm or 200 slpm are typical in this arrangement. Once the flow gets higher than that, the body size really jumps and this allows flows up to 2500 slpm to be controlled or 9000 slpm flows to be measured (without a control valve).

The gas itself is an important consideration. Are you flowing one of the simple, common gasses such as air, oxygen, nitrogen, helium, hydrogen, argon, CO or CO2? If so, a simple design of the MFC can be used. If your gas is very dangerous, you may be required to treat it differently. If it is susceptible to contamination, such as reacting with moisture, the interior of the MFC may be polished and the connections changed from screwed or compressed fitting to higher purity type connections such as VCR or VCO which utilize metal seals and are considered when high and ultra high purity gases are being used. If the process is being run at a positive pressure, the design of the MFC is such that it can contain that pressure. Some MFCs are limited to 100 psi or 500 psi maximum working pressures. Industrial MFCs are usually standard at 1500 psi (100 bar) and can be made to withstand 4500 psi (300 bar) pressure sometimes found in catalyst research, hydrogenation of food or drugs, or many petrochemical processes. On the other hand, if your process operates under vacuum, the design of the MFC has to be adjusted accordingly. Gases behave differently at positive pressures than they do under high vacuum and calibrating an MFC for vacuum service is a precise process. Temperatures of the gases are to be considered. Usually the gases are measured and controlled at relatively mild temperatures and elevated after the MFC if high temperatures are required in the process. There are temperature limits inside an MFC due to the measuring technology which utilizes the thermal properties of the gas to determine the flow. So, typically, 140-160°F is the working limit for gasses in MFCs.

When asked about accuracy, many applications require the most accurate system available. However some processes are not precise and a less accurate device can be used. Sometimes accuracy is not as important as repeatability. You must first understand the difference between percent of full scale accuracy statements and percent of rate accuracy statements. They mean the same thing only when the device is operating at 100 percent of its rated capacity. If you ask the MFC to control the gas flow to a lower set point, the amount of error varies in different MFC designs. Older, analog MFCs typically have accuracy statements such as +/- 1 % of full scale. This means a 100 sccm MFC would have an acceptable error of 1 sccm at any flow. If the flow were to be set at 50%, a 1% FS error would actually be 2% of that rate. If the flow were at 20%, the same 1 sccm deviation would now be a 5% error. Devices with error statements given in % fs are not as accurate as those given in % rate. Let us look at a device with a stated accuracy of +/- 1% of rate from 100% flow down to 20% flow and +/- 0.2% of full scale below 20 percent flow. Using the same example, if the 100 sccm MFC were flowing at 50% the maximum error would be ½ sccm or 1 % of rate. If the flow were set at 20%, the error would be 0.2 sccm or 1% of that rate. Please see Figure 1 which shows the error bands for % fs and % of rate.

accuracy comparison: percent full scale vs. percent rate 

Figure 1: Accuracy comparison: % full scale vx. % rate. 

Communicating electrically with an MFC can be done in a number of ways. The oldest and most basic would be to use an analog signal of 0-5 volts DC. 0 volts indicates no flow and 5 volts indicates the maximum flow. You will have to either search the calibration data to determine just what the maximum flow is for your particular MFC or, sometimes, it is printed on the body of the MFC. Any voltage between 0 and 5 volts would indicate a proportional flow somewhere in between zero and the maximum flow. 0-5 volt signals were very popular years ago and are still used today as an economical way to communicate with an MFC. Be aware, 0-5 volt signals do not travel well over long distances. It is common to limit them to 10 or 20 feet maximum. Another analog communication signal commonly used is 4-20 milliamps. A current of 4 mA on a 24 volt system indicates zero flow. Anything less than 4 mA would be abnormal and could be used to alarm. A current of 20 mA indicates full flow and 12 mA would be 50%. Any other currents between 4 and 20 mA would be proportional to the flow. Both current and voltage signals are still in use today for simplicity and economics. Once a digital circuit board is built in to the MFC, you have many options in the way of communications. The basic RS485 protocol can be used by many third party softwares. DeviceNet, ProfiBus, and Foundation FieldBus are very powerful communications systems that allow the user to not only to control and read the flow, but to perform many diagnostic and alarm subroutines all-the-while controlling and reading the flow of the gas.

At the heart of the most accurate MFCs is a Hastelloy® sensor to protect against aggressive gases. You will also find a feature called MultiFlo™ which overcomes a long-standing limitation of many thermal MFCs – when changing gas types, a simple correction factor, such as the ratio of heat capacities between the calibration gas and new gas, cannot account for accuracy-robbing viscosity and density differences. The Brooks MultiFlo™ database is built on thousands of native gas runs to establish correction functions that account for both thermal and physical differences among gases making the GF Series among the most accurate and flexible MFCs/MFMs available today. This type of MFC is the perfect choice for customers who use thermal mass flow controllers or thermal mass flow meters on a variety of gases, who need to change gas type frequently, or who need to re-range while preserving gas measurement and control accuracy.

Some examples:

  • OEMs will reduce the number of gas- and range-specific MFCs that they inventory.
  • Solar, biotech, CVD, plasma, glass, web coating, nanotechnology,vacuum processing and similar large users of mass flow meters and mass flow controllers will greatly reduce their gas- and range-specific spares inventory.
  • R&D, research, and laboratory users can quickly change experiment conditions and achieve much better actual process gas accuracy vs. traditional mass flow devices.

MultiFlo™ programming is simple and fast – a new gas and range can be programmed under 60 seconds plus the device can be programmed without removing it from service or disconnecting the device from any process or tool control system.



In summary, a mass flow meter or mass flow controller can be a very simple device to measure and control the flow of a gas or it can be utilized in the most demanding of applications; but only if you address the four major considerations detailed above (flow range, gas conditions, desired accuracy, and communications).